Program Overview
Aeris V500 · Critical-Care Ventilator · Stage: Verification
Live program telemetry
Aeris V500 · Verification phase
Waveform reflects live agent + human activity across the traceability spine
Requirements decomposed
428
+12 this week
Standards mapped
10
1 gap
Verification coverage
87%
+3.4pt
BOM saving / unit
$218
of $246 target
Total requirements
428
236 SW · 192 HWCritical-to-Safety
112
94% verifiedTest coverage
87%
1,284 casesBOM saving / unit
$218
12.4% of BOMVerification coverage trend
By test level · nightly rollup
Compliance readiness
Weighted alignment · monthly
Program workflow
Live pipeline status · entities in stage / total
Decompose
100%
Classify
96%
Map compliance
91%
Generate tests
84%
Optimize BOM
72%
Agent activity
Live · from audit + agent runs
streaming
- Requirement→Test Generator2 min agoGenerated 14 proposed test cases for HW-221 (blower drive)HW-221
- Verification Runner6 min agoTC-08840 apnea backup timing FAILED — 4.2s vs BC-06 threshold 4.0sREQ-1058
- L. Okafor · Compliance Lead18 min agoApproved 3 boundary conditions for IEC 81001-5-1
- CTS/CTQ Classifier (Beta)34 min agoProposed CTS classification for 7 requirements — awaiting review
- Alternate-Vendor Agent1 h agoIdentified qualified alternate for SEN-O2-G1 (City Tech → Sensirion)SEN-O2-G1
- M. Chen · Test Engineer2 h agoOverrode classification on REQ-0431 (CTQ → CTS) with rationale
Your approvals
Items awaiting sign-off
- Safety failure dispositionREQ-1058 · TC-08840Apnea backup 4.2s exceeds BC-06 4.0s threshold. Recommend retest with FW 2.3.1-rc3.
- Classification overrideREQ-0431Reclassify CTQ → CTS (Part 11 audit trail is safety-adjacent).
- Compliance mappingIEC 81001-5-1 → SW-160Auto-mapped 6 requirements under §5.2 patch SLA.
- Test case approvalHW-221 · 14 casesGenerated cases covering blower drive envelope + edge current profile.