Product Breakdown

Aeris V500 · decomposed to part level for safety-heavy subsystems

Total requirements
428
SW / HW split
236 / 192
Nodes
76
Deepest branch
L4
Decomposition tree
Click a node to view details · re-decompose scoped to any subtree
SYS
SYS-0Aeris V500 System428 reqL1
SW
SW-0Ventilation Software Stack236 reqL2
SW
SW-100Breath control loop
CTS
72 reqL3
SW
SW-120Alarms & monitoring
CTS
58 reqL3
SW
SW-140Clinician UI
CTQ
61 reqL3
SW
SW-160Cybersecurity services
CTS
45 reqL3
HW
HW-0Ventilator Hardware192 reqL2
HW
HW-200Pneumatic subsystem
CTS
84 reqL3
HW
HW-300Control PCBA
CTS
54 reqL3
HW
HW-400Enclosure & mounts
Standard
24 reqL3
HW
HW-500Battery pack
CTQ
30 reqL3
SYS
SYS-0L1
Aeris V500 System

Element responsible for aeris v500 system. 428 requirements resolve under this node.

Domain
SYS
Requirements
428
Classification
Governing std
ISO 80601-2-12
Version
v3.2
Last change
2h ago · R. Vasquez
Comments
R. Vasquez3d ago
Split further per DFMEA review 3/12 — need dedicated inspiratory valve node.