Verification (V-Model)
1,284 test cases · 87% coverage · live rollup from CI rig + manual entries
Overall coverage
All test cases · live
87%
1,284 cases
V-model traceability
Each definition level paired with its verification activity
| Definition | Verification activity | Cases | Pass | Fail | Running |
|---|---|---|---|---|---|
| User Needs | Clinical Validation | 42 | 38 | 0 | 4 |
| System Requirements | System Verification | 218 | 190 | 3 | 25 |
| Subsystem Design | Integration Test | 486 | 431 | 7 | 48 |
| Detailed Design | Unit Test | 538 | 471 | 11 | 56 |
SRS generated
82/96
85%
SRS reviewed
71/96
74%
SRS approved
62/96
65%
Recent test executions
Results from CI rig via ALM integration + manual entries
| Test case | Requirement | Name | Level | Result | When | By |
|---|---|---|---|---|---|---|
| TC-08840 | REQ-1058 | Apnea backup ventilation start ≤4.0s | System | Fail | 34 min ago | Verification Runner |
| TC-08834 | REQ-1042 | Blower pressure step response ±2 cmH2O | Integration | Pass | 58 min ago | CI rig |
| TC-08821 | REQ-3140 | PEEP stability across cycles | System | Running | 12 min ago | CI rig |
| TC-08810 | REQ-2145 | Trigger response ≤100 ms | Integration | Pass | 2 h ago | CI rig |
| TC-08801 | REQ-4020 | Humidifier reaches 37°C in 15 min | System | Pass | 3 h ago | M. Chen |