Traceability Explorer

One graph, one table — click REQ-1058 and see the unbroken chain

SW
HW
BC
REQ
TC
STD
containsclassifiesverified-byexercisessourced-fromSW-122Apnea detectionBC-06Apnea start ≤4.0sREQ-1058Backup ventilation startTC-08840Apnea backup testHW-221Blower driveISO 80601-2-12§201.12.1.103
Trace table · REQ-1058
HopFromRelationshipToDirectionStatus
1SW-122 Apnea detectioncontainsREQ-1058
upstream
active
2BC-06 Apnea start ≤4.0sclassifiesREQ-1058
upstream
active
3ISO 80601-2-12 §201.12.1.103sourced-fromBC-06
upstream
active
4REQ-1058verified-byTC-08840
downstream
Fail
5TC-08840exercisesHW-221 Blower drive
downstream
active