Compliance & Standards

10 standards mapped · 10 active boundary conditions

Mapped standards
StandardBodyCategoryCoverageRequirementsStatus
IEC 60601-1IEC
Safety
96%
118
Aligned
IEC 60601-1-2IEC
EMC
92%
41
Aligned
IEC 62304IEC
Software
91%
82
Aligned
IEC 81001-5-1IEC
Cyber
78%
26
Gap
ISO 14971ISO
Risk
94%
63
Aligned
ISO 13485ISO
Quality
89%
48
Partial
IEC 62366-1IEC
Usability
85%
22
Partial
ISO 80601-2-12ISO
Safety
93%
71
Aligned
FDA 21 CFR 820FDA
Quality
88%
44
Partial
MDR 2017/745EU
Regulatory
82%
39
Partial
Coverage by category
Current vs target readiness
Derived boundary conditions
Versioned · Compliance Lead approval required before applying to live requirements
IDParameterThresholdDrivesLinked reqsSource
BC-01Tidal volume accuracy±4%
CTS
12ISO 80601-2-12 §201.12.4.101
BC-02Peak pressure clamp≤ 60 cmH2O
CTS
8ISO 80601-2-12 §201.12.4.108
BC-03FiO2 delivery accuracy±3% abs
CTS
6ISO 80601-2-12 §201.12.4.103
BC-04Alarm audibility (bed)≥ 65 dBA @ 1m
CTQ
5IEC 60601-1-8
BC-05Battery runtime (transport)≥ 45 min
CTQ
4Benchmark (Hamilton C6)
BC-06Apnea backup ventilation start≤ 4.0 s
CTS
3ISO 80601-2-12 §201.12.1.103
BC-07PEEP stability window±1 cmH2O
CTQ
7Benchmark (Dräger Evita)
BC-08SW cybersecurity update SLA≤ 30 days
CTS
6IEC 81001-5-1 §5.2
BC-09Trigger response latency≤ 100 ms
CTS
5ISO 80601-2-12 §201.12.4.107
BC-10Humidifier reach temp≥ 37°C in 15 min
CTQ
4ISO 8185
Competitor benchmarking
Parameter-by-parameter vs named competitors and regulatory floor
ParameterAeris V500Hamilton C6Dräger EvitaReg. floorVerdict
Tidal volume accuracy±4%±5%±4%±10%
Leads
Apnea backup start≤4.0s≤4.5s≤5.0s≤5.0s
Leads
Battery runtime45 min60 min45 min30 min
On par
Trigger latency≤100 ms≤90 ms≤120 ms≤200 ms
Behind